Аналізатор заліза Liquiline System CA80FE, Endress+Hauser

Аналізатор заліза Liquiline System CA80FE, Endress+Hauser
Попереднє замовлення
Доступне передзамовлення
Endress+Hauser Виробник: Endress+Hauser
Без ПДВ: 0.00UAH

Colorimetric system for monitoring of drinking water, mineral water and process water

Field of application

Liquiline System CA80FE offers precise, regulation-compliant online measurement. It supports you in controlling the iron content of drinking water, mineral water or source water for industrial processes. Like all Liquiline System analyzers, it enables plug & play of up to four Memosens sensors – minimizing the installation effort. Automatic calibration and cleaning and the low reagent consumption reduce operating costs while advanced diagnostics with remote access ensure process safety.

Liquiline System CA80FE monitors:

  • Iron content in drinking water and mineral water

  • Residual iron after precipitation, flocculation and filtration processes

  • Process water containing diluted iron

Benefits

  • Compliance with standard ferrozine method ensures direct comparability to lab measurements

  • Supports predictive maintenance by advanced diagnostics

  • Reduced operating costs through automatic calibration and cleaning as well as low reagent consumption

  • Easy upgrade of functionality - even to a complete measuring station - simply by adding modules and connecting Memosens sensors.

  • Comfortable remote access via integrated web server

  • Seamless integration into process control systems thanks to digital field busses such as Modbus, PROFIBUS or EtherNet/IP

Specifications

Measuring principle

Colorimetric

Characteristic

Process analyzer for iron in aqueous solutions
Comply with standard colorimetric measuring principle, ferrozine method

Size

Housing (open version):
793 x 530 x 417 mm
31.22 x 20.87 x 16.42 in
Housing (closed version):
793 x 530 x 463 mm
31.22 x 20.87 x 18.23 in
Housing with base:
1723 x 530 x 463 mm
67.83 x 20.87 x 18.23 in

Design

Open design, cabinet and stand housing
High-Performance plastic ASA-PC, additional stand coated steel

Process temperature

4 to 40 °C (39.2 to 104 °F)

Ambient temperature

5 to 40 °C (41 to 104 °F)
Outdoor version: - 20 to 40 °C (-4 to 104 °F)

Process pressure

At atmospheric pressure, < 0.2 bar

Sample flow rate

Min. 5 ml/min (0.17 fl.oz/min)

Consistency of the sample

Suspended solids content
Turbitity < 50 NTU, aqueous, homogenized

Application

Monitoring of the iron content of potable water, mineral water and source water for industrial processes

Power supply

100 to 120 VAC / 200 to 240 VAC ± 10%
24 VDC ± 10%
50 ± 1 or 60 ± 1,2 Hz

Output / communication

2x 0/4 to 20 mA
Webserver, Modbus, Ethernet/IP, Profibus DP

Input

1 or 2 measuring channel
1 to 4 digital sensor inputs for sensors with Memosens protocol (optional)

Measurement range

0.05 to 2 mg/l Fe
0.1 to 5 mg/l Fe
0.1 to 5 mg/l with dilution function to maximum 2.0 to 100 mg/l Fe

Consumables

Reagents and standard solutions CY80FE for the operation
Regular maintenance is done with the parts of the maintenance kit CAV800