iTHERM MultiSens Flex TMS02 багатоточковий термометр RTD/TC, Endress+Hauser

iTHERM MultiSens Flex TMS02 багатоточковий термометр RTD/TC, Endress+Hauser
Попереднє замовлення
Доступне передзамовлення
Endress+Hauser Виробник: Endress+Hauser
Без ПДВ: 0.00UAH

Modular direct contact TC and RTD multipoint thermometer with secondary process barrier

Field of application

The iTHERM TMS02 MultiSens Flex can be customized for various applications: Bendable into a 3D layout, it provides a complete temperature profile using only a single process connection. The device is configurable with or without thermowells for optimal response times, mechanical strength or sensors replaceability. It is available with either TC/RTD technology or with our iTHERM StrongSens and features an advanced diagnostics chamber for additional safety.

  • Petrochemical industry

  • Heavy Chemical processes

Benefits

  • Infinite layouts of 3D sensors distribution for any desired process monitoring configuration

  • High degree of customization thanks to a modular product design for easy installation, process integration and maintenance

  • High degree of insert process compatibility as per standard IEC 60584, ASTM E230 and IEC 60751

  • On-board electronic heat protection for highest product durability

  • Compliance to different types of protection for use in hazardous locations for easy process integration

  • Diagnostic chamber as secondary process barrier for additional safety, capable of pressure monitoring for predictive maintenance through advanced diagnostics

Specifications

Measuring principle

Resistance Temperature Detector

Characteristic / Application

metric style
easy-to-use
suitable for hazardous areas
flanged process connection
light chemical processes
3D sensors distributions
modular design
replaceable sensors during shutdown
diagnostic chamber







Thermowell / protection tube

without (for direct contact sensors type)
with (for replaceable sensors)
with (for increased mechanical protection)
Multi-thermowells, each one dedicated to a sensor


Insert / probe

mineral insulated (MI), flexible
mineral insulated (MI), flexible, with its own thermowell

Outer diameter protection tube / Insert

6,0 mm (0,24'')
8,0 mm (0,31")
10,29 mm (1/8")

Max. immersion length on request

up to 30.000,00 mm (1181'')

Material protection tube/ thermowell

1.4401 (316)
1.4435 (316L)
1.4541 (321)
1.4550 (347)
Alloy 600 (2.4816)



Process connection

flange:
2" ASME/ANSI B16.5 150
...600
3" ASME/ANSI B16.5 150
..600
4" ASME/ANSI B16.5 150
...600
6" ASME/ANSI B16.5 150
...600
8" ASME/ANSI B16.5 150
...600
DN80 EN/DIN 1092.1 PN10
...PN100
DN100 EN/DIN 1092.1 PN10
...PN100
DN125 EN/DIN 1092.1 PN10
...PN100
DN150 EN/DIN 1092.1 PN10
...PN100
DN200 EN/DIN 1092.1 PN10
...PN100



Tip shape

straight
reinforced
easy-fix

Operating temperature range

Type K:
max. 1.150 °C
(max. 2.102 °F)
Type J:
max. 720 °C
(max. 1.328 °F)
Type N:
max. 1.100 °C
(max. 2.012 °F)
Pt100 WW; 3mm; 316L; -200...600oC
Pt100 WW; 6mm; 316L; -200...600oC
Pt100 TF; 3mm; 316L; -50…400oC
Pt100 TF; 6mm; 316L; -50…400oC
Pt100 StrongSens; 6mm; 316L; -50…500oC

Max. process pressure (static)

at 20 °C: 200 bar (2900 psi)

Accuracy

class 2 acc. to IEC 60584
ASTM E230 and ANSI MC 96.1
IEC/Class A
IEC/Class AA

Response time

depending on configuration:
TC:
t50 = 2 s
t90 = 5 s
RTD:
t50 = 0,8s
t90 = 2s

Integration head transmitter

yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

Ex - approvals

ATEX
IECEX
UL
FM
CSA
NEPSI
EAC
UK CA
CCC Ex ia IIC T1...T6 Ga

Certification

SIL (transmitter only)