TEC420 Термопарний термометр, Endress+Hauser

TEC420 Термопарний термометр, Endress+Hauser
Попереднє замовлення
Доступне передзамовлення
Endress+Hauser Виробник: Endress+Hauser
Без ПДВ: 0.00UAH

Comprehensive and common temperature measurement technology for almost all branches of industry

Field of application

The high modular and robust thermometer finds its use in several applications in almost all industries. An optional head transmitter, with all common communication protocols, makes the device ready to use with enhanced measurement accuracy and reliability compared to directly wired sensors. A variety of process connections, dimensions and materials offer flexible application possibilities.

TEC420 sheathed thermocouples cover temperature ranges from -40°C to 1100°C depending on the type of thermocouple. They can be used in low pressure applications in direct contact with the medium or protected by a thermowell. They can also be used in all applications where small immersion lengths are needed. Special attention should be paid to the maximum operating temperature of the connection head (80 °C if transmitter included).

    Benefits

    • High degree of flexibility thanks to modular design with standard terminal heads as per DIN EN 50446 and customer-specific immersion lengths

    • High degree of insert compatibility and design as per DIN 43772

    • Head transmitter with easy selection: Analog output 4 to 20 mA, HART®, PROFIBUS® PA or FOUNDATION Fieldbus™

    Specifications

    Measuring principle

    Thermocouple

    Characteristic / Application

    metric style
    modular temperature assembly
    suitable for hazardous areas
    process connection as compression fitting
    without neck

    Thermowell / protection tube

    without (not intended to use with thermowell)

    Insert / probe

    mineral insulated (MI), flexible

    Outer diameter protection tube / Insert

    6,0 mm (0,24'')

    Max. immersion length on request

    up to 100.000,00 mm (3.937,01'')

    Material protection tube/ thermowell

    1.4401 (316)
    Alloy 600 (2.4816)

    Process connection

    depending on configuration

    Operating temperature range

    -40 °C ...1.100 °C
    (-40 °F ...2.012 °F)

    Max. process pressure (static)

    at 20 °C: 20 bar (290 psi)

    Accuracy

    class 2 acc. to IEC 60584

    Response time

    t50 = 3 s
    t90 = 7 s

    Integration head transmitter

    yes (4 … 20 mA; HART; PROFIBUS PA; FOUNDATION FIELDBUS)

    Ex - approvals

    ATEX

    Certification

    SIL (transmitter only)